Pipette Calibration - Addendum The calibration process described in "Repair and Calibration of Pipettes" is incomplete. The pipette should be checked at all three points specified in the Rainin published specs. These can be found in the Pipetman instruction manual. Below are relevant excerpts. Model Volume(uL) Mean Error(uL) -------------------------------------- P-2 0.2 0.024 1.0 0.027 2.0 0.030 P-10 1 0.025 5 0.075 10 0.1 P-20 2 0.1 10 0.1 20 0.2 P-100 10 0.35 50 0.4 100 0.8 P-200 50 0.5 100 0.8 200 1.6 P-1000 100 3 500 4 1000 8 Often, the pipette will perform within the specified error window at one extreme of its range but not at the other. Sometimes compromises can be made in calibration to satisfy both extremes. For example: A P-1000 has been calibrated at the maximum volume and is measured at 999, well within the window of error. However, when checked at the 100 microliter setting it measures 104, outside the specified window. The volume is reduced slightly by adjusting the calibration wheel. Now the measurement at the maximum setting is 993, and at the 100 microliter setting it is 101. Both measurements are within the specified error windows. The next step would be to check the 500 microliter setting. Sometimes a very tricky 3-way compromise is needed. If a compromise between settings at the extremes of the pipette's range is not obtainable, but most of the pipette's range can be calibrated within spec, some users are satisfied to calibrate the portion of the range most used. For example, a P-1000 that can be calibrated from 200 to maximum is often considered usable. The user would switch to a calibrated P-200 for lower volumes. If the error measured at the first calibration point is very large, say 10% or 20% of full range, the instrument may need to be rezeroed. This could be the result of the threaded metering shaft having been completely removed for some reason. Bring the instrument to the repair shop for rezeroing.