Repair and Calibration of Pipettes - PAGE 2 c. Place the proper tip firmly on the pipette. Set the volume to the maximum capacity by turning the dial to about five units above the maximum then back down smoothly to the desired volume. This reduces error due to backlash in the dial and screw. Press the plunger down until a resistance is felt. Hold the plunger and immerse the tip in the DI water 1-2 mm for the smaller pipettes and 2-4 mm for the P1000. Slowly release the plunger while holding the pipette close to vertical. When the plunger is fully released, wait 2 seconds for the liquid to settle. Lift the pipette and check for leakage; if the water drips out, the seals need to be replaced. Check for any water clinging to the outer surface of the tip. If there is any, carefully wipe it off without disturbing the liquid inside. Check for air bubbles in the sample; if any, start over and release the plunger more slowly. d. Zero the balance with the weighing container. Dispense the water into the container slowly, then push the plunger past the resistance to expell all of the water. Check that no water remains in the tip. Cap the container and weigh. Repeat the procedure several times and compare the reproducibility of your results with the specifications in Table 1. The first sample may be different because the plastic tip absorbs some water at first. Don't try to change the calibration until you get reproducible readings. Erratic readings may be due to poor technique, bad seals, a corroded piston, or or other damaged parts. e. If the volume is consistently outside the specified limits, the calibration needs to be changed. The method used depends on the model. Older units have locking setscrews. Newer units have a plastic calibration screw on the plunger rod. Older models: Turn the volume knob until the three setscrews above the rubber ring are exposed. Do not touch the three setscrews below the rubber ring. Use the 2mm hex wrench to loosen the upper setscrews. Then use the special tool to turn the slot at the plunger shaft. The slot is turned clockwise to reduce the volume. Then lock the adjustment by tightening the three setscrews gently. Reset the volume knob and check the volume using the balance. Newer models: Use the special wire tool to turn the plastic screw with holes that surround the plunger rod. Insert the tool tip fully into one of the small holes and turn the tool so that it rotates the plastic screw. Turn clockwise to reduce the volume. The adjustment is self-locking. Reset the volume knob and check the volume using the balance. Other Problems: The following problems are more difficult to fix. Turn these in for repair. 1. If calibration is not possible within the adjustment range, the dial needs to reset. 2. Bent plunger rods can be straightened. 3. Cracked rubber friction rings can be replaced.